Development and calibration of high precision counter torque measurement system
DOI:
Author:
Affiliation:

Clc Number:

TN6;TH7

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    In response to the need for high-precision micro-torque measurement of the torque fluctuation of the gyroscope motor, a highprecision counter-torque measurement system based on a torque device and an angle sensor is proposed. Based on a coaxially transmitted stepped shaft structure, the external wiring is flexibly connected with spring like hairspring to reduce the elastic interference torque caused by wiring, and realize the high-precision measurement of the torque fluctuation; based on the traditional weight calibration, the static calibration and uncertainty analysis of the measurement system are carried out. The experimental results show that the system measurement accuracy is improved by 90% compared to before optimization. In the range of -10~ 10 mN·m, the measurement accuracy is as high as 0. 06%, the linearity is better than 0. 03%, and the drift error is better than 0. 5 μN·m/ 2 h, the uncertainty is better than 0. 025%.

    Reference
    Related
    Cited by
Get Citation
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online: February 27,2023
  • Published: