刘士兴,李江晖,夏 进,易茂祥,梁华国.应用于 ATE 的时间测量单元设计[J].电子测量与仪器学报,2023,37(6):86-92
应用于 ATE 的时间测量单元设计
Design and implementation of time measuring unit applied to ATE
  
DOI:
中文关键词:  时间测量单元  时间数字转换器  数字 IC 交流参数测量  集成电路自动测试设备
英文关键词:time measurement unit  time to digital converter  AC parameter measurement for digital IC  automatic test equipment for IC
基金项目:国家自然科学基金重大科研仪器研制项目(62027815)资助
作者单位
刘士兴 1.合肥工业大学微电子学院 
李江晖 1.合肥工业大学微电子学院 
夏 进 1.合肥工业大学微电子学院 
易茂祥 1.合肥工业大学微电子学院 
梁华国 1.合肥工业大学微电子学院 
AuthorInstitution
Liu Shixing 1.School of Microelectronics, Hefei University of Technology 
Li Jianghui 1.School of Microelectronics, Hefei University of Technology 
Xia Jin 1.School of Microelectronics, Hefei University of Technology 
Yi Maoxiang 1.School of Microelectronics, Hefei University of Technology 
Liang Huaguo 1.School of Microelectronics, Hefei University of Technology 
摘要点击次数: 244
全文下载次数: 590
中文摘要:
      集成电路飞速发展对集成电路自动测试设备(ATE)中时间测量单元(TMU)的精度提出了更高的要求。 针对这一问 题,本文使用电子学引脚测试芯片 MAX9979 对数字 IC 施加激励和捕获响应,结合 Xilinx Artix-7 FPGA 内部固化的时间数字转 换器(TDC)设计了一种高精度的时间测量单元。 时间数字转换器采用粗、细计数结合的内插方法,粗计数由参考时钟为 200 MHz 的 32 位直接计数器实现;细计数由超前快速进位链(CARRY4)级联的延迟链构成,通过对 CARRY4 进行专用配置来 减小其超前进位功能引起的测量误差,使用码密度校准法对延迟链进行校准。 实验结果表明,TMU 量程为 21. 475 s,平均分辨 率为 34. 7 ps,DNL 优于 2. 5 LSB,INL 优于 4. 5 LSB,精度为 39. 7 ps。
英文摘要:
      The accuracy of time measurement units ( TMU) in integrated circuit automatic test equipment ( ATE) is facing higher demands due to the rapid development of integrated circuits. To cope with the increased accuracy of digital IC time parameter measurements, a high precision time measuring unit was designed in this paper, which used the electronic pin test chip MAX9979 to apply excitation and capture response, and combined with Xilinx Artix-7 FPGA internal curing time digital converter (TDC). Time-todigital converter implemented by interpolation using a combination of coarse and fine counting, the coarse counter is implemented by a 32-bit direct counter with a reference clock of 200 MHz, while the fine counter consists of a delay chain cascaded by a CARRY4 cascade, with a dedicated configuration for CARRY4 to reduce the measurement error caused by its overrun function, and the delay chain was calibrated using the code density calibration method. The experimental results show that the TMU range is 21. 475 s, the average resolution is 34. 7 ps, the DNL is less than 2. 5 LSB, the INL is better than 4. 5 LSB, and the accuracy is 39. 7 ps.
查看全文  查看/发表评论  下载PDF阅读器