何梓滨,邢 昊,蒋廷勇,柴 昊,吴梦娟,王晓嘉,张荟如.超大型低反射镜面单锥脉冲电场标准装置[J].电子测量与仪器学报,2023,37(5):117-126
超大型低反射镜面单锥脉冲电场标准装置
Ultra large low reflection mirror monocone pulsed electric field standard device
  
DOI:
中文关键词:  镜面单锥  脉冲电场标准  阻抗匹配  电波暗室
英文关键词:mirror monocone  pulsed electric field standard  impedance matching  anechoic chamber
基金项目:航天科技集团计量科研项目资助
作者单位
何梓滨 1. 北京东方计量测试研究所 
邢 昊 1. 北京东方计量测试研究所 
蒋廷勇 2. 中国人民解放军 63660 部队 
柴 昊 1. 北京东方计量测试研究所 
吴梦娟 1. 北京东方计量测试研究所 
王晓嘉 2. 中国人民解放军 63660 部队 
张荟如 1. 北京东方计量测试研究所 
AuthorInstitution
He Zibin 1. Beijing Oriental Institute of Metrology and Test 
Xing Hao 1. Beijing Oriental Institute of Metrology and Test 
Jiang Tingyong 2. No. 63660 Unit of the Chinese People’s Liberation Army 
Chai Hao 1. Beijing Oriental Institute of Metrology and Test 
Wu Mengjuan 1. Beijing Oriental Institute of Metrology and Test 
Wang Xiaojia 2. No. 63660 Unit of the Chinese People’s Liberation Army 
Zhang Huiru 1. Beijing Oriental Institute of Metrology and Test 
摘要点击次数: 411
全文下载次数: 973
中文摘要:
      针对现有镜面单锥 TEM 室用作脉冲电场传感器校准时存在测试尺寸空间较小、低频响应较差、内部与环境反射影响测 量准确度等问题,采用超大型精密单锥结构、终端阻抗加载和低反射异形电波暗室等方法,在仿真优化分析基础上,设计研制了 基于超大型低反射镜面单锥的脉冲电场标准装置。 该镜面单锥母线长 2 m,能够适应脉冲宽度大于 5 ns 的脉冲信号,整体装置 所产生的标准脉冲电场幅值 50~ 150 V/ m 可调、脉冲前沿小于 100 ps、脉冲宽度大于 5 ns,满足装备测试脉冲电场传感器的大空 间、宽频带、高准确度校准需求。
英文摘要:
      In view of the problems of small test size space, poor low-frequency response, and internal and environmental reflection affecting measurement accuracy when the existing specular monocone TEM cell is used as the calibration of pulsed electric field sensors, the standard device of pulsed electric field based on ultra-large low-reflection specular monocone is designed and developed on the basis of simulation optimization analysis by using ultra-large precision single-cone structure, terminal impedance loading and low-reflection special-shaped anechoic chamber. The generatrix of the monocone is 2 m long and can adapt to pulse signals with a pulse width greater than 5 ns, and the amplitude of the standard pulse electric field generated by the device is 50~ 150 V/ m, the rise time of the pulse is less than 100 ps, the pulse width is greater than 5 ns. This standard device meets the needs of large space, wide frequency band and high accuracy calibration equipped with test pulse electric field sensors.